Data Acquisition
Data acquisition is the generation of data that can be manipulated with a computer.
CT Scanner
A computed tomography (CT) scanner is used to create cross-sectional slices of different objects. The medical version of CT system scans the human body for tumors or other abnormalities, other versions are used for non-destructive testing in the industry.
The CT imaging system includes the moveable gantry and patient table or couch. The gantry is a frame that contains the x-ray source, collimators, filters, detectors, a data acquisition system (DAS), rotational components including slip ring systems and all associated electronics. The x-ray tube and detector system are mounted opposite each other, allowing a rapid and synchronous rotation around the patient table.
In older CT scanners a small generator supplied power to the x-ray tube and the rotational components via cables for operation. Up to the 4th generation the CT tube and detectors rotate together around the patient for each slice. CT systems with slip ring technology (the x-ray tube rotates around a stationary ring of detectors) operate without cables and provides continuous rotation of the gantry components without interference of cables. Spiral CT scanners work with a continuous table movement while the x-ray tube is rotating around the patient.

Overview about CT scanner generations:
1st generation has a pencil beam and one detector;
2nd generation has a narrow fan beam and multiple detectors;
3rd generation has a wide fan beam;;
4th generation is equipped with a detector ring;
5th generation is the electron beam tomography CT
6th generation is the helical (or spiral) CT scanner;
7th generation is the multi-detector row CT.

See also Contrast Media Injector, Dual-Head CT Power Injector, Syringeless CT Power Injector.
ECG Trigger
ECG trigger methods synchronize the heartbeat in order to minimize motion artifacts, whereat the r-wave is used as the trigger. ECG gating techniques are useful whenever data acquisition is too slow to occur during a short fraction of the cardiac cycle.

See also Triggering and Trigger Delay.
Gantry
The gantry is a ring-shaped structure, containing the x-ray tube, collimators, filters, data acquisition system (DAS), associated electronics such as gantry angulation motors, rotational components including slip ring systems and the detector array in a CT or radiation therapy system. The table control, to regulate the gantry tilt and laser (or high intensity) lights are included within, or mounted on the gantry serving as anatomical positioning guides. To scan the patient, the patient table is moved through the gantry aperture.
The rotating arm on which an accelerator head is mounted is also called gantry. This gantry can rotate 360 degrees around its axis.
Trigger Delay
The trigger delay is the time after triggering at which data acquisition takes place.
Result Pages :
 
Share This Page
Facebook
Twitter
LinkedIn

Look
      Ups
Radiology - Technology Information Portal
Member of SoftWays' Medical Imaging Group - MR-TIP • Radiology-TIP • Medical-Ultrasound-Imaging
Copyright © 2008 - 2024 SoftWays. All rights reserved.
Terms of Use | Privacy Policy | Advertising
 [last update: 2023-11-06 02:01:00]