Radiology - Technology Information Portal
Thursday, 2 May 2024
• Welcome to Radiology-TIP.com!
     • Sign in / Create account
 
 'Field of View' p2
SEARCH   
A B C D E F G H I J K L M N O P Q R S T U V W X Y Z 
Searchterm 'Field of View' found in 4 terms [
] and 6 definitions [
], (+ 3 Boolean[
] results
Result Pages :
Contrast
Contrast is the relative difference of intensities in two adjacent regions of an image. When referring to computed tomography (CT), contrast is defined as a difference in Hounsfield units between structures. The measurement of contrast resolution in CT imaging involves determining how easy it is to differentiate tissues whose CT density is similar to that of their surroundings. An image lacks contrast when there are no sharp differences between black and white. Brightness refers to the overall lightness or darkness of an image.
The contrast between air, soft tissue, and bones in x-ray and CT images is based on their different absorption of x-rays. Differences in tissue density, thickness and changes of the x-ray spectrum have consequences for image contrast, image noise as well as patient dose.
Optimized tube current, collimation, pitch and image reconstruction improves the contrast. Higher image contrast is produced by increased slice thickness, smaller matrix, and large field of view which results in large voxel size; high mAs to reduce noise; low pass filter.

See also Contrast Enhanced Computed Tomography.
Display
A display is a computer monitor that shows the processed data from the scanned area. Displays can be black-and-white or color, small or large depending upon the model and price of the machine.

See also Display Field of View, Display Matrix, Window Width, Bit Range, Hounsfield Scale, Interpolation, Minimum Intensity Projection, and Printer.
Gantry Aperture
The gantry aperture is the opening through which the table with the patient moves for the examination. Typical diameters of the gantry range from 50 to 85 cm. The gantry aperture diameters between 70 and 85 cm are useful for biopsy procedures and obese patients. Larger gantry apertures allow for easier handling of the biopsy equipment.
The scanning field of view in which the CT data are acquired is smaller than the gantry aperture diameter.
Pixel
A pixel is a picture element (pix, abbreviation of pictures + element). Tomographic images are composed of several pixels; the pixel size is determined by the used field of view and the number of elements in the display image matrix. The corresponding size of the pixel may be smaller than the actual spatial resolution.
Pixels do not have a fixed size; their diameters are generally measured in micrometers (microns). Although the pixel is not a unit of measurement itself, pixels are often used to measure the resolution (or sharpness) of images. As a hypothetical example, a 600 x 1000 pixel image has 4 times the pixel density and is thus 4 times sharper than a 300 x 500 pixel image, assuming the two images have the same physical size.
Resolution
In medical imaging, resolution is the ability to distinguish two adjacent objects and a measurement of image quality. The resolution of tomographic images is a function of slice thickness, field of view (FOV) and matrix size. The resolution in plane is a function of FOV / matrix size.
Result Pages :
 
Share This Page
Facebook
Twitter
LinkedIn
Look
      Ups
Radiology - Technology Information Portal
Member of SoftWays' Medical Imaging Group - MR-TIP • Radiology-TIP • Medical-Ultrasound-Imaging
Copyright © 2008 - 2024 SoftWays. All rights reserved.
Terms of Use | Privacy Policy | Advertising
 [last update: 2023-11-06 02:01:00]