Radiology - Technology Information Portal
Tuesday, 23 April 2024
• Welcome to Radiology-TIP.com!
     • Sign in / Create account
 
 'Pitch Factor' 
SEARCH   
A B C D E F G H I J K L M N O P Q R S T U V W X Y Z 
Searchterm 'Pitch Factor' found in 1 term [
] and 1 definition [
]
Result Pages :
Pitch Factor
The pitch factor is a selectable CT parameter. The definition depending on the IEC standards is the ratio of the table speed per rotation and the total collimation.
The choice of pitch factor is determined on practical considerations and on the availability of the computed tomography algorithms and parameters for image reconstruction.

See Pitch.
Pitch
(p) The pitch (in computed tomography) is the ratio of the patient table increment to the total nominal beam width for the CT scan. The pitch factor relates the volume coverage speed to the thinnest sections that can be reconstructed. In spiral CT, dose is always inversely proportional to pitch.
Among the different manufacturers, there are various definitions of pitch depending on whether a single-detector (single-slice) or multi-detector (multi-slice) CT scanner is used.
For a single-slice helical scanner the pitch is:
Pitch = table movement per rotation/slice collimation.
The international standard formula (by the IEC) is:
Pitch = TF (table feed in mm per 360° rotation)/(N (number of detector rows) x SC (slice collimation in mm))
Result Pages :
 
Share This Page
Facebook
Twitter
LinkedIn

Look
      Ups
Radiology - Technology Information Portal
Member of SoftWays' Medical Imaging Group - MR-TIP • Radiology-TIP • Medical-Ultrasound-Imaging
Copyright © 2008 - 2024 SoftWays. All rights reserved.
Terms of Use | Privacy Policy | Advertising
 [last update: 2023-11-06 02:01:00]