Filtration
X-rays contain a range of energies (polychromatic photons), the higher energies pass through the patient, the lower energies are absorbed or scattered by the body. Ideally, the x-ray beam should be monochromatic or composed of photons having the same energy. Strong filtration of the beam results in more uniformity. The more uniform the beam, the more accurate the attenuation values or CT numbers are for the scanned anatomical region.
There are two types of filtration utilized in CT:
Inherent tube filtration and filters made of aluminum or Teflon are utilized to shape the beam intensity by filtering out the undesirable x-rays with low energy. Filtration of the x-ray beam is usually done by the manufacturer prior to installation. The half value layer provides information about the energy characteristics of the x-ray beam. Too much filtration produces a loss of contrast in the x-ray image.
A mathematical filter such as a bone or soft tissue algorithm is included into the CT reconstruction process to enhance resolution of a particular anatomical region of interest.
CT Scanner
A computed tomography (CT) scanner is used to create cross-sectional slices of different objects. The medical version of CT system scans the human body for tumors or other abnormalities, other versions are used for non-destructive testing in the industry.
The CT imaging system includes the moveable gantry and patient table or couch. The gantry is a frame that contains the x-ray source, collimators, filters, detectors, a data acquisition system (DAS), rotational components including slip ring systems and all associated electronics. The x-ray tube and detector system are mounted opposite each other, allowing a rapid and synchronous rotation around the patient table.
In older CT scanners a small generator supplied power to the x-ray tube and the rotational components via cables for operation. Up to the 4th generation the CT tube and detectors rotate together around the patient for each slice. CT systems with slip ring technology (the x-ray tube rotates around a stationary ring of detectors) operate without cables and provides continuous rotation of the gantry components without interference of cables. Spiral CT scanners work with a continuous table movement while the x-ray tube is rotating around the patient.

Overview about CT scanner generations:
1st generation has a pencil beam and one detector;
2nd generation has a narrow fan beam and multiple detectors;
3rd generation has a wide fan beam;;
4th generation is equipped with a detector ring;
5th generation is the electron beam tomography CT
6th generation is the helical (or spiral) CT scanner;
7th generation is the multi-detector row CT.

See also Contrast Media Injector, Dual-Head CT Power Injector, Syringeless CT Power Injector.
X-Ray Tube
X-ray tubes are devices for the production of x-rays. X-ray tubes consist of an evacuated glass vessel and two electrodes. An electrical current with very high voltage passes across the tube and accelerates electrons emitted by thermionic emission from a tungsten filament (cathode also called electron gun) towards the anode target. The electrons collide with the anode and this deceleration generates x-rays (bremsstrahlung).
The high vacuum allows the electron beam an unimpeded passage. The electron beam heats the anode (usually copper), which is cooled by water to prevent melting. A copper target emits x-rays with a characteristic wavelength. Other used metals soften or harden the x-ray beam.
The x-rays pass through a very thin beryllium (Be) foil. This beryllium window absorbs a high amount of the elastically scattered electrons (produced by the target) and allows the radiation to get out of the tube without substantial absorption.
In conventional x-ray tubes, the anode is also the target. In nanofocus and microfocus x-ray tubes, the electron beam is transmitted through a hole in the anode where it is then focused onto a small spot on the target.

See also X-Ray Tube Housing, Fine Focus X-Ray Tube, Transformer, Diode, Digital to Analog Converter and Angular Response.
Attenuation
The attenuation of radiation is a decrease in intensity as a result of interactions by transmission through matter. X-ray beams attenuate due to photon absorption by the material or scattering. Both effects are energy dependent. The probability of absorption or scattering is a function of the photon energy. The photoelectric absorption is much more energy dependent than the Compton scatter effect.

See also Attenuation Correction, Linear Energy Transfer, Broad Beam and Ion Beam.
Filter
Different filter types are used in medical imaging to improve image quality.
Filter materials such as aluminum, iron or copper, located in the primary x-ray beam between tube and collimator, filter out lower energetic x-rays (for example bow-tie filter). Filters partially absorb or attenuate the x-ray beam to prevent radiation overexposure. The sum of filtration consists of inherent and added filters. This filtration affects the beam energy and ability to penetrate materials.
Mathematical filter procedures are used for the convolution of the attenuation profiles and reconstruction of computed tomography (CT) images.
Result Pages :
 
Share This Page
Facebook
Twitter
LinkedIn
Look
      Ups
Radiology - Technology Information Portal
Member of SoftWays' Medical Imaging Group - MR-TIP • Radiology-TIP • Medical-Ultrasound-Imaging
Copyright © 2008 - 2024 SoftWays. All rights reserved.
Terms of Use | Privacy Policy | Advertising
 [last update: 2023-11-06 02:01:00]