Filter grids are used to reduce scattered noise and increase contrast in x-ray images. Primary radiation passing through an object gets scattered caused by the various density of different materials. Scatter radiation produces noise (radiographic fog) on the film or detector, which degrades the diagnostic quality. Anti-scatter grids act as filters between patient and film (or receiver) to remove scatter radiation. The use of a grid is recommended with body parts thicker than 10 cm and kVp values about 60 kV.
X-ray filter grids are available with focused or parallel strips. These two types are produced with linear or crossed grid configurations. The septa of filter grids consist of high radiation absorbing materials (e.g. lead) separated by permeable parts. During radiation exposure, movement of the grid blurs a projection of the septa.
If the image receptor and x-ray tube (with the focal spot) are in a fixed position relative to one another the grid is automatically aligned. In mobile radiography, the position of the focal spot and the image receptor is variable. Additionally cassettes incorporating anti-scatter grids are also available.
X-Ray Tube
X-ray tubes are devices for the production of x-rays. X-ray tubes consist of an evacuated glass vessel and two electrodes. An electrical current with very high voltage passes across the tube and accelerates electrons emitted by thermionic emission from a tungsten filament (cathode also called electron gun) towards the anode target. The electrons collide with the anode and this deceleration generates x-rays (bremsstrahlung).
The high vacuum allows the electron beam an unimpeded passage. The electron beam heats the anode (usually copper), which is cooled by water to prevent melting. A copper target emits x-rays with a characteristic wavelength. Other used metals soften or harden the x-ray beam.
The x-rays pass through a very thin beryllium (Be) foil. This beryllium window absorbs a high amount of the elastically scattered electrons (produced by the target) and allows the radiation to get out of the tube without substantial absorption.
In conventional x-ray tubes, the anode is also the target. In nanofocus and microfocus x-ray tubes, the electron beam is transmitted through a hole in the anode where it is then focused onto a small spot on the target.

See also X-Ray Tube Housing, Fine Focus X-Ray Tube, Transformer, Diode, Digital to Analog Converter and Angular Response.
Result Pages :
 
Share This Page
Facebook
Twitter
LinkedIn
Look
      Ups
Radiology - Technology Information Portal
Member of SoftWays' Medical Imaging Group - MR-TIP • Radiology-TIP • Medical-Ultrasound-Imaging
Copyright © 2008 - 2024 SoftWays. All rights reserved.
Terms of Use | Privacy Policy | Advertising
 [last update: 2023-11-06 02:01:00]