Filtration
X-rays contain a range of energies (polychromatic photons), the higher energies pass through the patient, the lower energies are absorbed or scattered by the body. Ideally, the x-ray beam should be monochromatic or composed of photons having the same energy. Strong filtration of the beam results in more uniformity. The more uniform the beam, the more accurate the attenuation values or CT numbers are for the scanned anatomical region.
There are two types of filtration utilized in CT:
Inherent tube filtration and filters made of aluminum or Teflon are utilized to shape the beam intensity by filtering out the undesirable x-rays with low energy. Filtration of the x-ray beam is usually done by the manufacturer prior to installation. The half value layer provides information about the energy characteristics of the x-ray beam. Too much filtration produces a loss of contrast in the x-ray image.
A mathematical filter such as a bone or soft tissue algorithm is included into the CT reconstruction process to enhance resolution of a particular anatomical region of interest.
Leakage Radiation
Leakage radiation is all radiation escaping from within the source assembly except for the useful beam. It is primarily controlled through the design of the tube housing and proper collimator filtering.
Raw Data
Raw data are the values of all measured detector signals during a scan. After calibration for fluctuations in tube output and beam hardening, the attenuation properties of each x-ray signal are accounted and correlated with the ray position. From these data the CT images are reconstructed including the use of mathematical procedures like convolution filtering and back projection.
Raw data can also be used for later reconstruction of additional planes and images by using a different convolution filter, zoom reconstruction, or an alternative CT number scale.

See also Magnification, Archiving, Number of Measurements and Convolution.
Result Pages :
 
Share This Page
Facebook
Twitter
LinkedIn

Look
      Ups
Radiology - Technology Information Portal
Member of SoftWays' Medical Imaging Group - MR-TIP • Radiology-TIP • Medical-Ultrasound-Imaging
Copyright © 2008 - 2024 SoftWays. All rights reserved.
Terms of Use | Privacy Policy | Advertising
 [last update: 2023-11-06 02:01:00]